29 Kinds of Respiratory Pathogens Combined Nucleic Acid

Short Description:

This kit is used for the qualitative detection of novel coronavirus (SARS-CoV-2), Influenza A virus (IFV A), Influenza B virus (IFV B), Respiratory syncytial virus (RSV), Adenovirus (Adv), human metapneumovirus (hMPV), rhinovirus (Rhv), Parainfluenza virus type I/II/III (PIVI/II/III), human bocavirus (HBoV), Enterovirus (EV), Coronavirus (CoV), Mycoplasma pneumoniae (MP), Chlamydia pneumoniae (Cpn), and Streptococcus pneumoniae (SP) and Influenza A virus subtype H1N1(2009)/H1/H3/H5/H7/H9/H10, Influenza B virus Yamagata/Victoria, Human coronavirus HCoV-229E/ HCoV-OC43/ HCoV-NL63/ HCoV-HKU1/ MERS-CoV/ SARS-CoV nucleic acids in human oropharyngeal swab and nasopharyngeal swab samples.


Product Detail

Product Tags

Product name

HWTS-RT160 -29 Kinds of Respiratory Pathogens Combined Nucleic Acid Detection Kit

Epidemiology

Respiratory tract infection is the most common disease in humans, which can occur in any gender, age and region. It is one of the leading causes of morbidity and mortality in the population worldwide[1]. Common respiratory pathogens include novel coronavirus, Influenza A virus, Influenza B virus, Respiratory syncytial virus, Adenovirus, human metapneumovirus, rhinovirus, Parainfluenza virus type I/II/III, Bocavirus, Enterovirus, Coronavirus, Mycoplasma pneumoniae, Chlamydia pneumoniae, and Streptococcus pneumoniae, etc[2,3]. The symptoms and signs caused by respiratory infections are relatively similar, but the treatment methods, efficacy and course of infections caused by different pathogens are different[4,5]. Currently, the main methods used in laboratories to detect the above-mentioned respiratory pathogens include: virus isolation, antigen detection and nucleic acid detection, etc. This kit detects and identifies specific viral nucleic acids in individuals with signs and symptoms of respiratory infection, with typing detection of influenza viruses and coronaviruses, and combines with other laboratory results to provide aid to the diagnosis of respiratory pathogen infections. Negative results do not exclude respiratory viral infection and should not be used as the sole basis for diagnosis, treatment, or other management decisions. A positive result cannot rule out the bacterial infections or mixed infections by other viruses that outside the test indicators. Experimental operators should have received professional training in gene amplification or molecular biology detection, and have relevant experimental operation qualifications. The laboratory should have reasonable biosafety prevention facilities and protection procedures.

Technical Parameters

Storage

-18℃

Shelf-life 9 months
Specimen Type Throat swab
Ct ≤38
CV <5.0%
LoD 200Copies/μL
Specificity The cross-reactivity test results showed that there was no cross reaction between this kit and Cytomegalovirus, Herpes simplex virus type 1, Varicella-zoster virus, Epstein-Barr virus, Pertussis, Corynebacterium, Escherichia coli, Haemophilus influenzae, Lactobacillus, Legionella pneumophila, Moraxella catarrhalis, attenuated strains of Mycobacterium tuberculosis, Neisseria meningitidis, Neisseria, Pseudomonas aeruginosa, Staphylococcus aureus, Staphylococcus epidermidis,  Streptococcus pyogenes,Streptococcus salivarius, Acinetobacter baumannii,Stenotrophomonas maltophilia, Burkholderia cepacia, Corynebacterium striatum, Nocardia, Serratia marcescens, Citrobacter, Cryptococcus, Aspergillus fumigatus, Aspergillus flavus, Pneumocystis jiroveci, Candida albicans, Rothia mucilaginosus, Streptococcus oralis, Klebsiella pneumoniae, Chlamydia psittaci, Coxiella burnetii, and human genomic nucleic acids.
Applicable Instruments Applied Biosystems 7500 Real-Time PCR Systems,

Applied Biosystems 7500 Fast Real-Time PCR Systems,

QuantStudio®5 Real-Time PCR Systems,

SLAN-96P Real-Time PCR Systems (Hongshi Medical Technology Co., Ltd.),

LightCycler®480 Real-Time PCR system,

LineGene 9600 Plus Real-Time PCR Detection Systems (FQD-96A, Hangzhou Bioer technology),

MA-6000 Real-Time Quantitative Thermal Cycler (Suzhou Molarray Co., Ltd.),

BioRad CFX96 Real-Time PCR System,

BioRad CFX Opus 96 Real-Time PCR System.

Work Flow

Macro & Micro-Test Viral DNA/RNA Kit (HWTS-3017) (which can be used with Macro & Micro-Test Automatic Nucleic Acid Extractor (HWTS-3006C, HWTS-3006B)), and Macro & Micro-Test Viral DNA/RNA Kit (HWTS-3017-8) (which can be used with EudemonTM AIO800 (HWTS-EQ007)) by Jiangsu Macro & Micro-Test Med-Tech Co., Ltd.

The extracted sample volume is 200μL and the recommended elution volume is 150μL.